Chapter 1: General Information
Environmental Conditions
17
Environmental Conditions
Overvoltage Category
This product is intended to be powered by MAINS that comply to Overvoltage Category
II, which is typical of cord- and- plug connected equipment.
Pollution Degree
The 80000 Series Oscilloscope may be operated in environments of Pollution Degree 2
(or Pollution Degree 1).
Pollution Degree Definitions
Pollution Degree 1: No pollution or only dry, non- conductive pollution occurs. The
pollution has no influence. Example: A clean room or climate controlled office
environment.
Pollution Degree 2. Normally only dry non- conductive pollution occurs. Occasionally a
temporary conductivity caused by condensation may occur. Example: General indoor
environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non- conductive pollution occurs
which becomes conductive due to condensation which is expected. Example: Sheltered
outdoor environment.
Measurement Category
Measurement Category
The 80000 Series oscilloscope is intended to be used for measurements in Measurement
Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly connected
to MAINS. Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS derived circuits. In the latter case, transient stresses are
variable; for that reason, the transient withstand capability of the equipment is made
known to the user.
Measurement category II is for measurements performed on circuits directly connected
to the low voltage installation. Examples are measurements on household appliances,
portable tools and similar equipment.
Measurement category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit- breakers, wiring, including
cables, bus- bars, junction boxes, switches, socket- outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary motors
with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-
voltage installation. Examples are electricity meters and measurements on primary
overcurrent protection devices and ripple control units.
Transient Withstand Capability
The maximum transient withstand capability of this instrument is not specified. Do not
apply more than 5 volts rms at the oscilloscope’s vertical input BNC connectors.
Содержание Infiniium 8000 Series
Страница 6: ...6...
Страница 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Страница 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Страница 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Страница 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Страница 142: ...Chapter 7 Replaceable Parts Exploded Views 142 Power Supply and PC Motherboard Without Option 017...
Страница 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Страница 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Страница 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Страница 158: ...158...
Страница 160: ...160 Index...