Performance Verification for N2830A-Series Probes
7
InfiniiMax III+ Series Probes User’s Guide
135
Performance Test Record
NOTE
The recommended test interval is one year/2000 hours.
Table 25
Test Information
Keysight Technologies
Keysight InfiniiMax III+ Series Probe
Model Number:
Tested by:
Serial Number:
Work Order Number:
Recommended next test date:
Date:
Table 26
Test Results
Test with Probe Heads
(only required to test one)
Test Limits
Result
Pass/Fail
Bandwidth Performance Test
N2836A
≥
13 GHz
DC Input Resistance Performance Test
N2836A
98 to 102 k
W
(differential mode)
49 to 51 k
W
(single-ended mode)
N5441A
98 to 102 k
W
(differential mode)
49 to 51 k
W
(single-ended mode)
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...