Keysight N7000/1/2/3A

InfiniiMax III+ Probes

Handling 

Guide

Содержание InfiniiMax III+ N7000

Страница 1: ...Keysight N7000 1 2 3A InfiniiMax III Probes Handling Guide ...

Страница 2: ...oftware described in this document are furnished under a license and may be used or copied only in accor dance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcon tract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defi...

Страница 3: ...a DUT 9 Soldering Tips 13 In this guide you ll learn how to properly handle the probe to prevent damage and maintain high performance For more safe handling information go to www keysight com and search for InfiniiMax III ESD Best Practices demo video CAUTION InfiniiMax I and II probe heads cannot be used with Infiniimax III probe amplifiers and InfiniiMax III probe heads cannot be used with Infin...

Страница 4: ...re being connected All work including connecting probe amplifiers to the oscilloscope should be performed at a static safe work station as shown in Figure 1 Figure 1 Static Safe Work Station Many scopes including Keysight s 90000X series have a front panel ground socket You can plug the wrist strap into the ground socket as seen in the following picture CAUTION Electrostatic discharge ESD can quic...

Страница 5: ...ombination Both types when used together provide a significant level of ESD protection Of the two only the table mat and wrist strap combination provides adequate ESD protection when used alone To ensure user safety the static safe accessories must provide at least 1 MΩ of isolation from ground Purchase acceptable ESD accessories from your local supplier WARNING These techniques for a static safe ...

Страница 6: ...Probe Head From the Amplifier CAUTION Avoid damaging the probe amplifier connection pins As shown in Figure 3 never bend the probe head in order to pop it loose from the amplifier Do not wiggle the probe head up and down or twist it to remove the connectors from the sockets CAUTION Never allow the probe head to be connected to the probe amplifier if the probe amplifier is not connected to the osci...

Страница 7: ...Safely Connecting the Probe to an Oscilloscope N7000 1 2 3A InfiniiMax III Handling Guide 7 Figure 4 Connecting the Probe to the Oscilloscope ...

Страница 8: ...be amplifier before disconnecting the probe amplifier from the oscilloscope moving the probe amplifier from one oscilloscope channel to another Figure 5 Probe Improperly Disconnected from Oscilloscope While Probe Head is Con nected to the Probe Amplifier CAUTION Never allow the probe head to be connected to the probe amplifier if the probe amplifier is not connected to the oscilloscope channel ...

Страница 9: ...coiling by introducing a very slight twist with each coil This allows the cable rotations to lie flat against each other and will eliminate the net twisting of the cable in the end CAUTION Probes are sensitive devices and should be treated with care Do not bend or kink the probe amplifier cable Do not drop heavy objects on the probe drop the probe from large heights spill liquids on the probe etc ...

Страница 10: ... protect both your equipment and designs from damage Although tack putty and low temperature hot glue are recommended you can also use other methods such as tape or hook and loop strain relief Keep in mind that different accessories have different cable stiffness You should choose a strain relief method appropriate for the cable stiffness For instance it is best to secure the stiffer N5439A near t...

Страница 11: ...y to the underside of the probe amplifier body and attach it to a rigid body near your DUT Figure 6 Probe Secured Using Tack Putty You can also use putty with a positioner such as the N2787A as shown in the following figure Figure 7 Using Putty With the N2787A 3D Probe Positioner The same positioner can also be used to support your probe amplifier ...

Страница 12: ... temperature hot glue to secure cables Figure 8 Probe Secured Using Low Temperature Hot Glue CAUTION Only use low temperature hot glue To remove the hot glue warm it with a heat gun set on low Only heat the hot glue enough to remove it CAUTION Do not use the hot glue on the probe head PC board or tip leads ...

Страница 13: ...ering iron station if possible Set the temperature of the soldering iron s tip to between 370 C and 420 C for non RoHS standards Use the smallest tip possible Use an optical aid of some sort microscope preferred Employ minimal dwell times on the solder joint 2 seconds Solder only the tip of the wire onto your DUT The solder should not get close to the existing solder ball on the tip Use enough flu...

Страница 14: ...14 N7000 1 2 3A InfiniiMax III Handling Guide Soldering Tips ...

Страница 15: ......

Страница 16: ...This information is subject to change without notice Keysight Technologies Inc 2015 Printed in USA First Edition February 2015 N7000 97000 N7000 97000 www keysight com ...