Keysight E8357D/67D & E8663D PSG User’s Guide 295
Peripheral Devices
N5102A Digital Signal Interface Module
This error is reported when the output FIFO is
overflowing in the
digital module. This error can be generated if an
external clock
or its reference is not set up properly, or if the
internal VCO is
unlocked.
806
Digital module output FIFO underflow
error; There are not enough samples
being produced for the current clock
rate. Verify that the digital module
clock is set up properly.
This error is reported when the output FIFO is
underflowing in the digital module. This error can
be generated if an external clock or its reference is
not set up properly, or if the internal VCO is
unlocked.
2.
If the port configuration is parallel or parallel interleaved, using an IQ
signal type, press the
Clocks Per Sample
softkey.
Notice that multiple clocks per sample can be selected. Some DACs
require the ability to clock multiple times for each sample; having a clocks
per sample value greater than one reduces the sample rate by a factor
equal to the selected number of clocks per sample. The sample rate is
viewed on the first-level and Data Setup softkey menus.
3.
Select the clocks per sample value to fit the test.
4.
Press the
Clock Source
softkey.
From this menu, select the clock signal source. With each selection, the
clock routing display in the signal generator clock setup menu will change
to reflect the current clock source. This will be indicated by a change in
the graphic.
5.
Select the clock source.
If External or Device is Selected
Press the
Clock Rate
softkey and enter the clock rate of the externally
applied clock signal.
NOTE
The clock phase and clock skew may need to be adjusted each time the clock
rate setting is changed. Refer to
Содержание E8663D
Страница 4: ...4...
Страница 58: ...44 Keysight E8357D 67D E8663D PSG User s Guide Signal Generator Overview Rear Panel...
Страница 186: ...172 Keysight E8357D 67D E8663D PSG User s Guide Optimizing Performance Wideband IQ FM Mode 3 2 GHz Option 018...
Страница 198: ...184 Keysight E8357D 67D E8663D PSG User s Guide Analog Modulation Configuring the LF Output Option UNT...
Страница 212: ...198 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Working with Filters...
Страница 226: ...212 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Configuring Hardware...
Страница 286: ...272 Keysight E8357D 67D E8663D PSG User s Guide AWGN Waveform Generator Configuring the AWGN Generator...
Страница 326: ...312 Keysight E8357D 67D E8663D PSG User s Guide Peripheral Devices Millimeter Wave Source Modules...
Страница 346: ...332 Keysight E8357D 67D E8663D PSG User s Guide Troubleshooting Returning a Signal Generator to Keysight Technologies...