![Keysight E6650A EXF Скачать руководство пользователя страница 439](http://html1.mh-extra.com/html/keysight/e6650a-exf/e6650a-exf_user-programming-manual_1961647439.webp)
6 List Sequence Measurements
Meas Setup
If the number of array values is less than the total number of analysis steps, the array will be auto repeated to fill in
all the analysis steps.
Example
:LSEQ:ACQ2:AST3:SET:TIME:OFFS 10ms
:LSEQ:ACQ2:AST3:SET:TIME:OFFS?
:LSEQ:ACQ2:LIST:SET:TIME:OFFS 500us, 1ms, 2ms
:LSEQ:ACQ2:LIST:SET:TIME:OFFS?
:LSEQ:NUMB:ACQ 2
:LSEQ:ACQ1:NUMB:AST 5
:LSEQ:ACQ2:NUMB:AST 5
:LSEQ:LIST:SET:TIME:OFFS 0.1s,0.2s,0.3s
:LSEQ:LIST:SET:TIME:OFFS? => 0.1, 0.2, 0.3, 0.1, 0.2, 0.3, 0.1, 0.2, 0.3, 0.1
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
Preset
0 s
State Saved
Saved in instrument state.
Range
0 s to 1 s
Min
0 s
Max
1 s
Initial S/W
Revision
A.05.00
Help Map ID
35133
Analysis Interval
This parameter sets the period over which measurements are made.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup, Analysis Setup
Remote
Command
-To set Analysis Interval individually:
[:SENSe]:LSEQuencer:ACQuire{1:512}:ASTep{1:1000}:SETup:TIME:INTerval <time>
[:SENSe]:LSEQuencer:ACQuire{1:512}:ASTep{1:1000}:SETup:TIME:INTerval?
-To set the Analysis Interval as an array:
[:SENSe]:LSEQuencer:ACQuire{1:512}:LIST:SETup:TIME:INTerval
<time>,<time>,<time>,….
[:SENSe]:LSEQuencer:ACQuire{1:512}:LIST:SETup:TIME:INTerval?
Sequence Analyzer User's & Programmer's Reference
439
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...