6 List Sequence Measurements
List Sequence Measurements
Command to setup multiple analysis steps
The RUI command is provided to facilitate user to generate a list of regular analysis steps for specified
acquisition. All the input parameters of the command are listed below:
•
stepNumber: specifies the number of analysis steps to be generated
•
analysisOffset: specifies the analysis offset of the first analysis step. Along with the step separation, it is
used to derive the analysis offset for the rest of analysis steps
•
analysisInterval: specifies the analysis interval for each analysis step
•
stepSeparation: sets the interval of slot to measure
•
startPower: specifies the expected power of the first analysis step. Along with the power step, it is used
to derive the expected power for the rest of analysis steps
•
powerStep: sets the power step size
•
measBitMap: specifies the measurement bit map for each analysis step
•
append: If ON, the new generated analysis steps will be appended at the end of the specified acquisition,
otherwise the existing analysis steps will be overwritten by the new generated analysis steps
parameter_
table_
23.52941
52.94118
Key Path
SCPI Only
Mode
Sequence Analyzer
Remote
Command
[:SENSe]:LSEQuencer:ACQuire{1:1000}:MASTeps:SETup
<stepNumber>,<analysisOffset>,<analysisInterval>,<stepSeperation>,< startPower
>,< powerStep >,< measBitMap >,<append>
Example
LSEQuencer:ACQuire1:MASTeps:SETup 40, 25us, 300us, 666.67us, 20dbm, –1, 1, OFF
State Saved
Saved in instrument state.
Initial S/W
Revision
A.12.00
Help Map
ID
0
Cooperation Parameter Enable
When Previously, as Acq Integration is not normal, for example, Acq Integration for Acq 1 is Primary, Acq
Integration of Acq 2 is Lower and Acq Integration of Acq 3 is Upper, then peak power and expected power
for Acq 2 and Acq 3 will be same as Acq 1 and can not be changed. In order to increase dynamic range for
some cases, peak power and expected power for each acquisition are supposed to be set independently as
Acq Integration is not normal, this parameter is used to enable this capability.
parameter_table_
23.52941
52.94118
308
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...