6 List Sequence Measurements
List Sequence Measurements
3. Declared function result for the 2nd specified channel [4 byte float]
…
(m + 1). Declared function result for the last (mth) specified channel [4 byte float]
ADC Over Range
1. ADC over-range occurred (1: true, 0: false) [2 byte short]
Spectrum Data
1. Number of points in the spectrum data, k [4 byte int]
2. Start frequency of spectrum data (Hz) [8 byte double]
3. Step frequency of spectrum data (Hz) [8 byte double]
4. FFT bin at 1st point (dBm) [4 byte float]
5. FFT bin at 2nd point (dBm) [4 byte float]
…
(k + 3). FFT bin at last (kth) point (dBm) [4 byte float]
Initial S/W
Revision
A.14.00
Format Data: Numeric Data (Remote Command Only)
This command specifies the format of the trace data input and output. It specifies the formats used for
trace data during data transfer across any remote port. It affects only the data format for setting and
querying trace data for the :TRACe[:DATA], TRACe[:DATA]?, :CALCulate:DATA[n]? and FETCh:SANalyzer
[n]? commands and queries.
Remote Command
:FORMat[:TRACe][:DATA] ASCii|INTeger,32|REAL,32 |REAL,64
:FORMat[:TRACe][:DATA]?
Notes
The query response is:
ASCii: ASC,8
REAL,32: REAL,32
REAL,64: REAL,64
INTeger,32: INT,32
When the numeric data format is REAL or ASCii, data is output in the current Y Axis unit. When the
data format is INTeger, data is output in units of m dBm (.001 dBm).
The INT,32 format returns binary 32-bit integer values in internal units (m dBm), in a definite length
block.
Dependencies
Sending a data format spec with an invalid number (for example, INT,48) generates no error. The
analyzer simply uses the default (8 for ASCii, 32 for INTeger, 32 for REAL).
Sending data to the analyzer which does not conform to the current FORMat specified, results in an
error. Sending ASCII data when a definite block is expected generates message –161 "Invalid Block
Data" and sending a definite block when ASCII data is expected generates message –121 "Invalid
Character in Number".
Preset
ASCii
Backwards Compatibility
Note that the INT,32 format is only applicable to the command, TRACe:DATA. This preserves
Sequence Analyzer User's & Programmer's Reference
293
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...