6 List Sequence Measurements
List Sequence Measurements
Notes
Option FP2 is required.
See notes for Fast Power Fetch for return format.
Initial S/W Revision
A.14.00
Binary Read Fast Power Measurement (Remote Command Only)
This command query is used as shorthand for an INIT command immediately followed by a FETC?
command. The returned results are in a binary format.
Mode
All
Remote
Command
:CALCulate:FPOWer:POWer[1,2,…,999]:READ?
:CALCulate:FPOWer:POWer[1,2,…,999]:READ1?
Example
:CALC:FPOW:POW1:READ?
:CALC:FPOW:POW1:READ1?
Notes
Option FP2 is required.
Returns m 4 byte floating point binary values (Little-Endian), where m corresponds to the number of
bandwidths defined.
Initial S/W
Revision
A.14.00
Diagnostic Binary Read Fast Power Measurement (Remote Command Only)
This command query is used as shorthand for an INIT command immediately followed by a FETC?
command. The returned results are in a binary format. This command is used primarily for diagnostic
purposes to test for ADC overloads and to visibly inspect the spectrum.
Mode
All
Remote
Command
:CALCulate:FPOWer:POWer[1,2,…,999]:READ2?
Example
:CALC:FPOW:POW1:READ2?
Notes
Option FP2 is required.
Note: Spectrum data is only returned if the IncludePowerSpectrum parameter is set to True. If
IncludePowerSpectrum is False, the number of spectrum points will be zero (0).
Units of the returned values are dependent on the Function parameter per channel (e.g. dBm for
BandPower, Hz for PeakFrequency).
Returns binary data (Little-Endian) that contains information on m amount of channels, along with ADC
over range and full spectrum data.
The following is the binary format of the response.
Bandwidth Return Value
1. Number of channels specified, m [4 byte int]
2. Declared function result for the 1st specified channel [4 byte float]
292
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...