Tutorial
7
Keysight E3633A and E3634A User’s Guide
187
The output will also change due to changes in the load. As the load increases the
output current will cause a small drop in the output voltage of the power supply
due to the output impedance R. Any resistance in the connecting wire will add to
this resistance and increase the voltage drop. Using the largest possible hook up
wire will minimize the voltage drop. Using the remote sense leads at the load will
compensate for lead resistance in the load leads.
Figure 7-6
Simplified diagram of common mode and normal mode sources of
noise
When the load changes very rapidly, as when a relay contact is closed, the
inductance in the hook up wire and in the power supply output will cause a spike
to appear at the load. The spike is a function of the rate of change of the load
current. When very rapid changes in load are expected, a capacitor with a low
series resistance, in parallel with the power supply, and close to the load is the
best way to minimize these voltage spikes.
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...