1
Calibration Procedures
22
Keysight E3632A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available, use
the accuracy requirements shown to select substitute calibration standards.
Table 1-1
Recommended test equipment
Instrument
Requirements
Recommended model
Test function
GPIB controller
Full GPIB capabilities
Keysight 82341C Interface
card
Programming and readback
accuracy.
Oscilloscope
100 MHz with 20 MHz
bandwidth
Keysight 54602B
Display transient response and
ripple and noise waveform.
RMS voltmeter
20 MHz
Measure rms ripple and noise.
Digital voltmeter
Resolution: 0.1 mV
Accuracy: 0.01%
Keysight 34401A
Measure DC voltages.
Electronic load
Voltage Range: 50 Vdc
Current Range: 10 Adc
Open and Short Switches
Transient On/Off
Keysight 6063B
Measure load and line
regulations and transient
response time.
Resistive loads (R
L
)
2.1
Ω
, 200 W
7.5
Ω
, 200 W
Measure ripple and noise.
Current monitoring resistor
(shunt)
0.01
Ω
, 0.01%
Constant current test setup.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply Service Guide ...
Страница 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 68: ...2 Service 70 Keysight E3632A Service Guide ...
Страница 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...
Страница 71: ...Keysight E3632A Service Guide 73 Component locator diagram for the main board assembly top serial MY53xx6xxx ...
Страница 72: ...74 Keysight E3632A Service Guide Component locator diagram for the front panel top serial MY53xx6xxx ...