4
Theory of Operation
102
Keysight E3631A Service Guide
A-to-D Converter
The analog-to-digital converter (ADC) is used to change dc voltages into digital
information. The circuitry consists of an integrator amplifier (U6 and U8), current
steering switch U5, resistors (R5, R4, R15 and R16), voltage reference (U3A and
U3B), ADC controller U16, and residue ADC U17 on the top board.
The ADC method used by the Keysight E3631A is called multislope III. Multislope
III is a charge balancing continuously integrating analog-to-digital converter. The
input voltage continuously forces charge onto the integrator capacitors C18 and
C19 through R15 on the top board.
Switch U5 steers fixed positive or negative reference currents onto the integrator
capacitors to cancel, or balance the accumulated input charge. The level shifted
(R19 and R20 on the top board) output of the integrator is checked every 2.66 ?A
by the U16 COMP input. Logic state machines in U16 control the U5 current
steering to continuously seek an approximate 2.5 V level on the integrator
amplifier output, FLASH. If the ADC input voltage ADC REF is between ±15 V, the
integrator output (FLASH) will remain within the 0 V to 5 V range of the U17
on-chip ADC. An input greater than +15 V may cause the integrator output in U8
at pin 6 (U8-6) to saturate at about -15 V. An input less than -15 V may cause U8
to saturate with an output of about +15 V. The U17 ADC input (FLASH) is clamped
to 0 V or 5 V by R21 and CR3 to protect U17.
The integrator amplifier is formed by U6 and U8. Resistors R10 and R11 affect the
amplifier stability. Amplifier oscillation may occur if their values are incorrect.
Amplifier U6 improves the offset voltage characteristics of integrator amplifier U8
on the top board.
Each analog-to-digital conversion occurs continuously. The ADC starts by
clearing the integrator slope count in U16. At the end of the integration period,
the slope count is latched. The slope count provides the most significant bits of
the input voltage conversion. The least significant bits are converted by the
on-chip ADC of U17.
Содержание E3631A
Страница 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Страница 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 32: ...1 Specifications 30 Keysight E3631A Service Guide Figure 1 1 Dimensions of Keysight E3631A power supply ...
Страница 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Страница 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Страница 98: ...3 Calibration Procedures 96 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Страница 110: ...4 Theory of Operation 106 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 124: ...6 Component Drawings 120 Keysight E3631A Service Guide E3631 60002 Component Locator for the Main Board Top ...
Страница 125: ...Component Drawings 6 Keysight E3631A Service Guide 121 E3631 60003 Component Locator for the Front Panel ...
Страница 126: ...6 Component Drawings 122 Keysight E3631A Service Guide E3631 60004 Component Locator for the Main Board Bottom ...
Страница 127: ...Component Drawings 6 Keysight E3631A Service Guide 123 Component Locator for the Main Board Top serial MY53xx6xxx ...
Страница 128: ...6 Component Drawings 124 Keysight E3631A Service Guide Component Locator for the Front Panel serial MY53xx6xxx ...
Страница 129: ...Component Drawings 6 Keysight E3631A Service Guide 125 Component Locator for the Main Board Bottom serial MY53xx6xxx ...
Страница 130: ...6 Component Drawings 126 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...