Settings
Available Key Settings
Description
Trigger Source
Key, IO or Rmt
Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO and Rmt.
Key (List Run/Stop Key)
Selects the Run Stopped softkey as a trigger source.
IO (DIO Trigger In)
Selects any configured digital IO (with Trigger Input
function) as a trigger source
Rmt (Remote Command)
Selects a remote interface command as a trigger
source.
Trigger Delay
0 to 3600 s
Sets the trigger delay in seconds.
Default is 0 s.
Repeat Count
1 to 256
Sets the list repeat count. This sets the number of times that a list is executed before it
completes.
Default is 1.
Continuous
-
Enables check box to repeat the list continuously.
Step 3 – Run the output sequence list
When Output LIST is initiated, all the List properties cannot be configured, which includes adding or removing
step from the list.
When Step mode transient system is initiated, properties such as voltage mode, current mode, trigger source
and trigger delay cannot be configured.
If both Voltage and Current Mode is set to Fix and Trigger Source is set to Key, pressing
Run
will automatically
set both voltage and current mode to List and start the list operation.
– Set Voltage Mode to List: Press
V Mode Lst
.
– Set Current Mode to List: Press
I Mode Lst
.
– Set Trigger Source to Key. Press
Trig Src Key
.
– Press the color-coded
[On]
key to enable the selected output.
– Press
Run
to start the list operation. To abort the operation, press
Stopped
.
– Press
Back
to exit and return to the previous menu.
78
Keysight E36300 Series User's Guide
Содержание E36300 Series
Страница 1: ...User s Guide E36300 Series Programmable DC Power Supplies ...
Страница 12: ...12 Keysight E36300 Series User s Guide ...
Страница 29: ...Front panel Rear panel E36312A and E36313A only Keysight E36300 Series User s Guide 29 ...
Страница 106: ...106 Keysight E36300 Series User s Guide ...