Step 4 – Use the All Outputs On and Off keys:
Once output delays have been set, press
[All On/Off]
to start the On delay and Off delay sequence.
[All On/Off] will turn ALL outputs on or off, whether they are configured to participate in an output on/off
delay sequence or not.
From the remote interface:
To program turn-on and turn-off delays for outputs 1 through 3:
OUTP:DEL:RISE 0.01,(@1)
OUTP:DEL:RISE 0.02,(@2)
OUTP:DEL:RISE 0.03,(@3)
OUTP:DEL:FALL 0.04,(@1)
OUTP:DEL:FALL 0.03,(@2)
OUTP:DEL:FALL 0.02,(@3)
To only include outputs 1 and 2 in a sequence :
OUTP:COUP:CHAN CH1,CH2
To turn on two coupled outputs in a sequence:
OUTP ON (@1:2)
64
Keysight E36300 Series User's Guide