Configure the inhibit input mode:
1. Press the
Output Settings
>
Output Inhibit
to configure the inhibit input mode.
2. Select the desired mode (
Off
,
Latched
or
Live
).
To clear the Inhibit protection function, first remove the external Inhibit signal. Then select
Source Settings
>
Protection Clear
for all outputs. This clears the Inhibit protection function and returns the output to its previous
operating state.
From the remote interface:
To select the Inhibit function:
DIG:PIN3:FUNC INH
To select the pin polarity:
DIG:PIN3:POL POS
DIG:PIN3:POL NEG
To set Inhibit mode to Latching:
OUTP:INH:MODE LATC
To set Inhibit mode to Live:
OUTP:INH:MODE LIVE
To disable the Inhibit signal:
OUTP:INH:MODE OFF
Fault/inhibit system protection
The following figure illustrates some ways that you can connect the Fault/Inhibit pins of the connector.
Keysight E36200 Series User's Guide
83
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...