Keysight E1412A Digital Multimeter User Guide
lvii
Triggering the Multimeter
Keysight E1412A Multimeter Application Information
Checking the Trigger Source
The
TRIGger:SOURce?
command returns
“BUS”
,
“EXT”
,
“IMM”
, or
“TTLT
n
”
to
show the present trigger source. The string is sent to the output buffer.
External Triggering
Use
TRIGger:SOURce EXTernal
to set the trigger source to external.
–
The trigger signal must be a low-true pulse with a pulse width greater than
1
s. The trigger signal level accepted is TTL (+5V maximum
negative-going to 0V). See the following diagram for the “Trig” input
requirement. The diagram also shows the “VM Complete” output you can
use to synchronize with a switch module.
–
The multimeter takes one reading (or the number specified by
SAMPle:COUNt
) for each external trigger received on the front panel “Trig”
BNC connector.
Internal Triggering
The trigger signal is always present in the internal triggering mode. This mode is
selected with the
TRIGger:SOURce IMMediate
command.
A CONFigure or MEASure? command automatically sets the
trigger source to IMMediate. You must follow the CONFigure
command with the TRIG:SOUR command to set the trigger source
to BUS, EXTernal or to TTLTrg<n>. The MEAS? command always
uses TRIG:SOUR IMM.