Keysight E1412A Digital Multimeter User Guide
Keysight E1412A Multimeter Application Information
Measurement Configuration
–
You cannot control the reading rate for ac measurements with integration
time because integration time is fixed at 10 PLCs for all ac measurements.
You must use a trigger delay to pace ac voltage and ac current
measurements.
–
NPLCs are not applicable to the
FREQuency
and
PERiod
functions.
Frequency and period measurements set resolution by specifying aperture
time. The aperture time for the
FREQuency
and
PERiod
functions default to
100mS. Specify an aperture time of 10mS for
4½-digits, 100mS for 5½-digits or 1 second for 6½-digits of resolution.
–
Set integration time using the following commands:
[SENSe:]<
function
>:NPLC <
number
>
(NPLC
s are not applicable for the
FREQ
and
PER
functions
)
[SENSe:]<
function
>:APER <
seconds
>
Autozero
Autozero applies to dc voltage, dc current and 2-wire resistance measurements.
The multimeter internally disconnects the input signal following each
measurement and takes a
zero reading
when autozero is enabled. Autozero
enabled is the default setting. It then subtracts the zero reading from the
preceding reading. This prevents offset voltages present on the multimeter's
input circuitry from affecting measurement accuracy.
–
When autozero is disabled (
OFF
), the multimeter takes one zero reading
and subtracts it from all subsequent measurements. It takes a new zero
reading each time you change function, range or integration time. You can
disable autozero on dc voltage, dc current and 2-wire ohms measurements
only (it is always disabled for ACV and ACI functions). Autozero is always
enabled when you select 4-wire ohms or ratio measurements.
–
The autozero mode is stored in volatile memory. The multimeter
automatically enables autozero at power-on and after a module reset.
–
Use the following command to disable autozero or select the
ONCE
parameter. The
OFF
and
ONCE
parameters have a similar effect. Autozero
OFF
does not perform a new zero measurement. Autozero
ONCE
performs
an immediate zero measurement.
[SENSe:]ZERO:AUTO OFF|ONCE|ON
Ranging
You can let the multimeter automatically select the range using autoranging or
you can specify a range. If you specify an expected value for the signal you are
measuring, the multimeter selects the range to accommodate the expected
input signal and turns autoranging off. Specify a range for faster measurements
to eliminate the autoranging time.