Keysight E1412A Digital Multimeter User Guide
xxxvii
Resistance Measurements
Keysight E1412A Multimeter Application Information
The following program shows SCPI examples used to store a
NULL
value.
CONF:RES
Set to 2-wire ohms function.
Short the lead resistance at the DUT location.
READ?
Measure the 2-wire ohms lead resistance.
Enter lead resistance value into computer.
CALCulate:FUNCtion NULL
Set math operation to NULL.
CALCulate:STATe ON
Turn math operation ON.
CALCulate:NULL:OFFSet <
value
>
Store the NULL offset value.
Subsequent 2-wire ohms measurements will subtract the null offset value from
the measurement thereby removing the lead resistance from the measurement.
The second way to store the 2-wire lead resistance as the
NULL
offset value is to
let the multimeter automatically do this with the first measurement. The first
measurement made after
CALCulate
function is set to
NULL
and the
STATe
is set
to
ON
stores the measured value as the null offset.
CONF:RES
Set to 2-wire ohms function.
Short the lead resistance at the DUT location.
CALCulate:FUNCtion NULL
Set math operation to NULL.
CALCulate:STATe ON
Turn math operation ON.
READ?
Measure the 2-wire ohms lead resistance.
Enter lead resistance value into computer. The value is automatically
stored in the multimeter's null offset register.
Remove the short from the lead resistance at the DUT location
and connect leads to your DUT.
READ?
Make a 2-wire ohms resistance measurement.
Enter lead resistance value into computer. The NULL value is
subtracted from the measurement to more accurately provide
the DUT resistance.
Power Dissipation Effects
When measuring resistors designed for temperature measurements (or other
resistive devices with large temperature coefficients), be aware that the
multimeter will dissipate some power in the device-under-test. If power
dissipation is a problem, you should select the multimeter's next higher
measurement range to reduce the errors to acceptable levels. Table 4 shows
several examples.
Table 2-3
DUT Power Dissipation
Range
Test Current
DUTPower at Full Scale
100
1mA
100
W
1k
1mA
1mW
10k
100
A
100
W
100k
10
A
10
W
1M
5
A
25
W
10M
500nA
2.5
W