Keysight E1412A Digital Multimeter
User Guide
ccxxiii
B
Keysight E1412A Multimeter Error
Messages
The following sections describe the types of errors the E1412A reports; Execution
Errors, Self-Test Errors and Calibration Errors. The error code is given (e.g., -101)
followed by the associated error message and a description of what the error
message means.
Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command header or within a parameter.
Example:
CONF:VOLT#DC
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a blank
space before or after a colon in the command header, or before a comma.
Example:
SAMP:COUN ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used a
comma instead of a colon, semicolon, or blank space – or you may have used a
blank space instead of a comma. Example:
TRIG:COUN,1
or
CONF:FREQ 1000 0.1
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
Example:
TRIG:COUN '150'
or
TRIG:COUN A
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may have
entered an extra parameter, or you added a parameter to a command that does
not accept a parameter. Example:
READ? 10