Keysight E1412A Digital Multimeter User Guide
Keysight E1412A Multimeter Module Setup
Using This Chapter
–
Resolution is from 4½-digits for fast measurements to 6½-digits for more
accuracy. Resolution is set by specifying the integration time in number of
power line cycles (NPLCs) or corresponding aperture time. Table 1 shows
the correlation between NPLCs and resolution.
Setting the Module Address Switch
The logical address switch factory setting is 24. Valid addresses are from
1 to 254 for static configuration (the address you set on the switch) and address
255 for dynamic configuration. The E1412A supports dynamic configuration of
the address. This means the address is set programmatically by the resource
manager when it encounters a module with address 255 that supports dynamic
configuration.
If you install more than one multimeter, each module must have a different
logical address. If you use a VXIbus command module, the logical address must
be a multiple of eight (e.g., 32, 40, 48, etc.) Each instrument must have a unique
secondary address which is the logical address divided by eight.
Table 1-1
Resolution of Power Line Cycles
Power Line Cycles
Resolution
0.02
0.0001 x Full-Scale
0.2
0.00001 x Full-Scale
1
0.000003 x Full-Scale
10
0.000001 x Full-Scale
100
0.0000003 x Full-Scale
When using an E1405A/B or E1406A as the VXIbus resource
manager with SCPI commands, the multimeter's address switch
value must be a multiple of 8