34
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Verification Tests
Performance Verification
Optional AC Voltage Gain Verification
These tests are NOT intended to be performed with every calibration. This
procedure is used to verify additional instrument AC voltage specifications.
Connect the AC Standard across the multimeter’s HI and LO input terminals as
shown in Figure 4-5.
1
Set up the multimeter as follows:
Reset multimeter
*RST
Configure for AC voltage
CONF:VOLT:AC
Select Slow AC filter
DET:BAND 3
2
Set the AC Standard controls as follows:
Output
1 V
Frequency
20 Hz
3
Set the multimeter to the 100 mV range using the following command:
Set range
VOLT:RANG .1
4
Perform the measurement using the READ? command.
5
Verify the result is within specified limits and record the result.
6
Change amplitude range using VOLT:AC:RANG <range> and make a
measurement for each ACV input and range shown in Table 4-6 verifying the
result is within specified limits. Record the result.
Table 4-7
AC Voltage Gain Verification Tests
INPUT
INPUT
FREQUENCY
Keysight E1412A
Range
Error from nominal
24 hour
90 day
1 year
1 V
20 Hz
1 V
± 600 μV
± 800 μV
± 900 μV
1 V
1 kHz
1 V
± 600 μV
± 800 μV
± 900 μV
1 V
20 kHz
1 V
± 600 μV
± 800 μV
± 900 μV
1 V
50 kHz
1 V
± 1.4 mV
± 1.6 mV
± 1.7 mV
1 V
100 kHz
1 V
± 6.3 mV
± 6.8 mV
± 6.8 mV
1 V
300 kHz
1 V
± 55 mV
± 55 mV
± 55 mV
10 V
1 kHz
10 V
± 6 mV
± 8 mV
± 9 mV
1 V
1 kHz
10 V
± 2.4 mV
± 3.5 mV
± 3.6 mV
100 mV
1 kHz
10 V
± 12.04 mV
± 13.05 mV
± 13.06 mV
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Страница 22: ...16 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Operating Instructions Warnings and Cautions ...
Страница 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Страница 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Страница 117: ...110 E1412A 6 1 2 Digit Multimeter Service Guide Keysight E1412A Multimeter Error Messages Execution Errors ...
Страница 118: ......