Keysight E1412A 6 1/2 Digit Multimeter Service Guide
33
Performance Verification
Verification Tests
Frequency Gain Verification Test Procedure
Q = Quick performance verification test points.
Connect the AC Standard across the multimeter’s HI and LO input terminals as
shown in Figure 4-7.
1
Set up the Keysight E1412A as follows:
Reset multimeter
*RST
Configure for frequency
CONF:FREQ
2
Set the AC Standard as follows:
Output
10 mV
Frequency
100 Hz
3
Set the multimeter to the 100 mVac range as follows: Set amplitude range,
100 mV FREQ:VOLT:RANG 0.1.
4
Perform the measurement using the READ? command.
5
Verify the result is within specified limits and record the result.
6
Change amplitude range using FREQ:VOLT:RANG <range> and make a
measurement for the 1V, 100 kHz input shown in Table 4-5 verifying the result
is within specified limits. Record the result.
Table 4-6
Frequency Verification Tests
INPUT
FREQUENCY
INPUT Amplitude
using a coaxial
input cable
QUICK
VERIFI-CATI
ON
Keysight
E1412A
Range
Error from nominal
24 hour
90 day
1 year
100 Hz
10 mV
100 mV
± 0.06 Hz
± 0.1 Hz
± 0.1 Hz
100 kHz
1 V
Q
1 V
± 6 Hz
± 10 Hz
± 10 Hz
Figure 4-7
Frequency Test Setup
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Страница 22: ...16 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Operating Instructions Warnings and Cautions ...
Страница 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Страница 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Страница 117: ...110 E1412A 6 1 2 Digit Multimeter Service Guide Keysight E1412A Multimeter Error Messages Execution Errors ...
Страница 118: ......