Horizontal Controls
2
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
45
The area of the normal display that is expanded is outlined with a box and the rest
of the normal display is ghosted. The box shows the portion of the normal sweep
that is expanded in the lower half.
To change the time/div for the Zoom window, turn the horizontal scale (sweep
speed) knob. As you turn the knob, the zoomed window time/div is highlighted in
the status line above the waveform display area. The Horizontal scale (sweep
speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left-to-right position of the
zoom window. The delay value, which is the time displayed relative to the trigger
point) is momentarily displayed in the upper-right portion of the display when the
delay time (
) knob is turned.
Negative delay values indicate you're looking at a portion of the waveform before
the trigger event, and positive values indicate you're looking at the waveform after
the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn the
horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to
"To isolate an event for frequency
To change the horizontal scale knob's coarse/fine adjustment
setting
1
Push the horizontal scale knob to toggle between fine and coarse adjustment
of the horizontal scale.
When
Fine
is enabled, turning the horizontal scale knob changes the time/div
(displayed in the status line at the top of the display) in smaller increments. The
time/div remains fully calibrated when
Fine
is on.
When
Fine
is turned off, the Horizontal scale knob changes the time/div setting in
a 1-2-5 step sequence.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...