272
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
24
Reference
CSV and ASCII XY file structure
In CSV or ASCII XY format the
Length
control selects the number of points per
segment. All segments are contained in the CSV file or in each ASCII XY data file.
For example: If the Length control is set to 1000 points, there will be 1000 points
(rows in the spreadsheet) per segment. When saving all segments there are three
header rows, so the data for the first segment starts at row 4. The second
segment's data starts at row 1004. The time column shows the time since the
trigger on the first segment. The top row shows the selected number of points per
segment.
BIN files are a more efficient data transfer format than CSV or ASCII XY. Use this
file format for fastest data transfer.
Minimum and Maximum Values in CSV Files
If you are running a Minimum or Maximum measurement, the minimum and
maximum values shown in the measurement display may not appear in the CSV
file.
Explanation:
When the oscilloscope's sample rate is 2 GSa/s, a sample will be taken every
500 ps. If the horizontal scale is set to 10 us/div, there will be 100 us of data
displayed (because there are ten divisions across the screen). To find the total
number of samples the oscilloscope will take:
100 us x 2 GSa/s = 200K samples
The oscilloscope is required to display those 200K samples using 640 pixel
columns. The oscilloscope will decimate the 200K samples to 640 pixel columns,
and this decimation keeps track of the min and max values of all the points that
are represented by any given column. Those min and max values will be displayed
in that screen column.
A similar process is used to reduce the acquired data to produce a record usable
for various analysis needs such as measurements and CSV data. This analysis
record (or
measurement record
) is much larger than 640 and may in fact contain
up to 65536 points. Still, once the # of acquired points > 65536, some form of
decimation is required. The decimator used to produce a CSV record is configured
to provide a best-estimate of all the samples that each point in the record
represents. Therefore, the min and max values may not appear in the CSV file.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...