Waveform Generator
19
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
207
•
High-Z
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
To use waveform generator logic presets
With logic level presets, you can easily set the output voltage to TTL, CMOS
(5.0V), CMOS (3.3V), CMOS (2.5V), or ECL compatible Low and High levels.
1
If the Wave Gen menu is not currently displayed on the oscilloscope's softkeys,
press the
[Wave Gen]
key.
2
In the Wave Gen menu, press the
Settings
softkey.
3
In the Wave Gen Settings menu, press the
Logic Presets
softkey.
4
In the Wave Gen Presets menu, press one of the softkeys to set the generated
signal's Low and High voltages to logic compatible levels:
To add noise to the waveform generator output
1
If the Wave Gen menu is not currently displayed on the oscilloscope's softkeys,
press the
[Wave Gen]
key.
Softkey (logic levels)
Low level
High level, 50 ohm
expected output load
High level, high-Z
expected output load
TTL
0 V
+2.5 V (TTL compatible)
+5 V
CMOS (5.0V)
0 V
Not Available
+5 V
CMOS (3.3V)
0 V
+2.5 V (CMOS
compatible)
+3.3 V
CMOS (2.5V)
0 V
+2.5 V
+2.5 V
ECL
-1.7 V
-0.8 V (ECL compatible)
-0.9 V
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...