Trigger Mode/Coupling
12
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
127
2
In the Trigger menu, press the
Holdoff
softkey; then, turn the Entry knob to
increase or decrease the trigger holdoff time.
Trigger Holdoff
Operating Hints
The correct holdoff setting is typically slightly less than one repetition of the
waveform. Set the holdoff to this time to generate a unique trigger point for a
repetitive waveform.
Changing the time base settings does not affect the trigger holdoff time.
With Keysight's MegaZoom technology, you can press
[Stop]
, then pan and zoom
through the data to find where the waveform repeats. Measure this time using
cursors; then, set the holdoff.
External Trigger Input
The external trigger input can be used as a source in several of the trigger types.
On 4-channel oscilloscopes, the external trigger BNC input is on the rear panel
and is labeled
EXT TRIG
.
On 2-channel oscilloscopes, the external trigger BNC input is on the front panel
and is labeled
Ext Trig
.
The external trigger input impedance is 1M Ohm. This lets you use passive probes
for general-purpose measurements. The higher impedance minimizes the loading
effect of the oscilloscope on the device under test.
To set the external trigger input options:
1
Access the External Trigger menu:
•
On 4-channel oscilloscopes:
i
Press the
[Trigger]
key in the Trigger section of the front panel.
ii
In the Trigger menu, press the
External
softkey.
•
On 2-channel oscilloscopes, press the
[External]
key in the Vertical section of
the front panel.
CAUTION
Maximum voltage at oscilloscope external trigger input
30 Vrms, 40 Vpk
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...