Keysight B2980B User’s Guide, Edition 1
239
Function Details
Trigger System
Trigger Source
B2980 supports the following trigger sources. The trigger source must be set to
the ARM event and the TRIGGER event individually.
•
AUTO (automatic internal, AINT): Trigger source best suited for the present
operating mode is automatically selected by the internal algorithms.
•
BUS: Remote interface trigger command such as GET, TRIGger, and *TRG
•
TIMER (timer, TIMer): Signal internally generated every interval set by the
Period field
•
INT1 or INT2 (internal): Signal from the internal bus 1 or 2
•
EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, or EXT7 (external): Signal from the DIO
pin
n
, which is an I/O port of the Digital I/O D-sub connector on the rear panel.
n
= 1 to 7
•
LAN: LXI trigger
•
TIN: Signal from the Trigger Input connector
Device Actions
B2980 supports the following device actions. Also see
•
Transient device action (voltage source output for B2985B/B2987B)
Voltage source channel starts voltage output for applying one output value if
the TRIGGER event was detected and the Source delay time elapsed.
•
Acquire device action (measurement)
Measurement channel starts measurement for obtaining one measurement
data if the TRIGGER event was detected and the Measure delay time elapsed.
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...