Preparing for Imaging
5
Keysight 5500 SPM User’s Guide
5-12
Figure 4-13
Move the lever to open the nose assembly disk.
4
Place the probe under the copper-colored spring clip on the nose
assembly disk. Use the alignment guides in the fixture to help locate
the probe laterally.
5
A small alignment spot on the fixture (
indicates the proper location for the cantilever tip. Place the probe
such that the tip is close as possible to this spot.
6
Move the lever to the left to close the nose assembly disk.
7
Use the tweezers to finely adjust the probe such that the cantilever is
aligned over the alignment spot. Only grasp the probe from the sides
to avoid damaging the cantilever.
8
Grasping the nose assembly disk from the edges, remove it from the
fixture and align it on the nose assembly body already in the scanner.
Inserting the Scanner and Connecting Cables
At this point the probe, nose assembly and scanner should all be
assembled into one unit.
1
Make sure there is adequate clearance below the scanner socket in
the middle of the microscope.
2
Place the scanner assembly into the scanner socket, with the
scanner’s frosted screen facing up and forward (
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...