Temperature Control
5
Keysight 5500 SPM User’s Guide
5-22
11
On the Lakeshore controller press
Heater Range
and select
Low
.
12
Press
the
Setpoint
button and enter the desired final temperature.
13
Allow the temperature to stabilize.
14
Initiate an approach.
Tips for Temperature Controlled Imaging
•
Make sure there is good thermal contact between the sample and the
sample plate. If possible, mount the sample using the liquid cell even
for ambient imaging.
•
Double-sided tape reduces thermal conductivity as well as
introducing sample drift. Therefore it should not be used to mount
samples for temperature controlled experiments.
•
It is possible to ramp the temperature while imaging. Use slow
ramps, typically less than 1
C per minute.
•
Every sample will react differently to temperature control. A thin
piece of graphite is a good test sample to use while setting up the
temperature control system.
•
Temperature fluctuations due to excessive gains will cause the
surface to appear wavy. Reduce gains on the Lakeshore controller to
reduce this waviness.
•
Imaging in the environmental chamber is recommended as it will
help keep the temperature stable. Purging the environmental cell
with a dry gas such as nitrogen will help control the sample
environment if condensation from cooling to below the dew point
becomes an issue.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...