Temperature Control
5
Keysight 5500 SPM User’s Guide
5-20
Note that Cable 3 includes a tab on the black-wire side of the connector
(
). The tab must face the Lo jack on the controller.
Figure 13-23
Tab side of Cable 3 must be inserted into the Lo jack
Water Cooling
When a sample is cooled using the Peltier sample plate the opposite side
of the Peltier device becomes hot. The hot side is water cooled to
decrease the minimum sample temperature, reduce power requirements
and prevent overheating. Inlets are provided on the underside of the
Peltier sample plate to connect water cooling tubing.
A gravity-fed water-cooling system (
mechanical pumping because it reduces the potential for vibration that
can affect imaging. Two reservoirs are provided with the Peltier sample
plate, one to be used as a source and the other as a receptacle to store
water for recycling. A height difference of three feet between the source
and receptacle gives a minimum temperature of -25
C or
approximately 50
C below room temperature. Ice can be added to the
reservoir to increase efficiency for a minimum temperature of -30
C.
CAUTION
Connect the water-cooling system and test for leaks
before
making any
electronic connections or turning on any components.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...