Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-3
X and Y Sensor Calibration
Initial calibration of the X and Y sensors consists of matching the ±10 V
output range of the sensors with the actual piezo travel, which varies
from scanner to scanner. No sample is required for this procedure.
1
Without a sample in place, load a cantilever and use the laser
alignment controls and camera window to align the laser and
detector mode to set up the microscope for
as described in
.
2
Choose
Controls > Setup > Controller
to confirm the appropriate
controller definition file.
3
Choose
Controls > Setup > Microscope
to confirm the appropriate
microscope definition file.
4
Verify that the correct scanner calibration file is selected in the
Scanner menu.
5
Choose
Scanner > Edit
to open the Scanner Setup dialog box. Select
the
Closed Loop
tab.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...