Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-18
all four measurements. An example of one of the corner angle
measurements is shown below in
.
Figure 8-16
Angle measurement after calibrating Non-orthogonality
Z Calibration
Sensitivity
After the X and Y dimensions have been calibrated, obtain an image of a
Z calibration standard and render the data as Tilted. Zoom in on a single
pit to minimize distortion. Position the cross section tool over a Z
feature. Place markers at the top and bottom of the feature. Measure the
step size.
Figure 8-17
Cross-section checking feature height for Z sensitivity
It is important to correctly position the markers for the Z feature
measurement. Place them in the center of the scan range. It may also be
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...