Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-41
Volume Spectroscopy
Volume Spectroscopy is also an extension of Force Spectroscopy. The
polarity and strength of the interaction between the AFM tip and the
sample is measured in user defined areas. Math plug-ins are used to
extract parameters such as stiffness and adhesion to produce a two
dimensional map of material properties. As in FlexGrid Spectroscopy, a
standard topography image is collected and the user selects a particular
volume of the image to analyze.
The procedure for Volume Spectroscopy is the same as for FlexGrid
Spectroscopy, but the user defines a volume to analyze, rather than
individual points. The resolution of the measurements is defined by the
volume of analysis, with up to image pixel resolution points. The
procedure described below covers general guidelines for Volume
Spectroscopy in Contact Mode. It is a user-dependent choice to select
Contact vs AC Mode.
Volume Spectroscopy Procedure for Contact Mode
This procedure is done after the system has been prepared for Contact
Mode imaging as described in
For more accurate measurements, use a Closed Loop scanner and follow
the CL scanner calibration prior to acquiring Force vs. Distance data.
Select the
CL Sweep
check box in the Setup tab of the Spectroscopy
window.
1
Collect a topographic image as described in
2
Choose
Mode > Contact
.
3
), set the following parameters
as shown below:
a
Set Gains as needed for the sample.
b
Set the Setpoint to
0 V
.
c
Maximize the Range by entering a very large number (ie. 999)
and PicoView will adjust to the maximum value for the scanner.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...