Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-16
12
In the Scan and Motor window’s
Scan
tab, enter:
a
Scan Speed of
1-2
ln/s.
b
Resolution of
256
.
c
Scan
Size
(in microns).
d
X Offset
and/or
Y Offset
values to set the center of the scan.
13
In the Realtime Images
window, choose to display
Topography
,
CSAFM/BNC Aux
(Phase) and
HEB Aux
(Amplitude). If using a
MAC III controller, select
Aux 1
and
Aux 2
.
14
In the Scan and Motor window, click the
Up
blue arrow to initiate a
scan starting at the bottom of the grid. Click the
Down
blue arrow to
initiate the scan from the top down. The image maps will begin to be
rendered in the Realtime images window.
Force Modulation Microscopy (FMM)
Force Modulation is another derivative of Contact Mode, with
similarities to AC Mode as well. In FM Mode, an additional 20-50 kHz
oscillation is applied to the cantilever. The amplitude and phase of
oscillation will change depending upon the modulus of the surface at
any given point.
The multi-purpose scanner and AAC nose assembly are used for FM
Mode. A specific Force Modulation cantilever is available through
Keysight; however, the best choice of cantilever is often experimentally
determined. A MAC Mode or MAC III controller is also required.
With the MAC Mode controller the following cables must be added:
•
Connect a BNC cable from the
Phase
output of the MAC Mode
controller to the
Aux In
of AFM Controller.
•
Connect a BNC cable from the
Amplitude
output of the MAC Mode
controller to the
Aux
BNC on the Head Electronics Box.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...