Final Sensitivity Calculation
To calculate the final sensitivity, use this equation:
For example, if the Gain switch is set to 10 and the module has a
10 nA/V preamp, the final sensitivity will be 1 nA/V.
STM Imaging Procedure
The procedure for STM imaging is as follows:
1
If you are using the multi-purpose scanner, insert the nose assembly
into the scanner (see
2
Insert a tip into the nose assembly or scanner. Grasp the tip with a
tweezers, then insert it into the hollow tube until it protrudes
approximately 2 mm (
Figure 7-6
Inserting STM tip wire in scanner
3
Place the scanner in the scanner bracket on the microscope.
4
Prepare the sample and place it on a sample plate. The sample must
be electrically isolated from the sample plate. The particular
mounting arrangement will depend on the sample type and size.
5
Attach an electrode from the sample plate to the sample. Lift the
spring-loaded electrode clip on the sample plate and insert the
Final Sensitivity
PreampSensitivity
Gain
=
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...