AC Modes
5
Keysight 5500 SPM User’s Guide
5-13
the tip. The standard sample plate with the magnet in the center for
holding sample pucks cannot be used for Top MAC imaging due to the
interference from the magnet in the sample plate.
Figure 6-9
Top MAC nose assembly
Q Control
An oscillating cantilever in AC Mode is influenced by complex
interaction forces between it and the surface. By carefully setting system
parameters the system can be made to operate entirely in the regime of
net-attractive forces, thereby reducing the effect of the probe tip on the
sample. The range in which the parameters have to be adjusted can be
narrow, however, making it difficult to maintain in real operation.
Q Control is a method that reduces damping of the system, increasing
the quality factor of the oscillating cantilever. Enhanced resonance
allows imaging with very low force and high phase sensitivity. The
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...