Preparing for Imaging
5
Keysight 5500 SPM User’s Guide
5-29
Using the Video System
The Keysight 5500 SPM includes a USB-based video system for
viewing the tip and sample. The optics of the video system and optics in
the scanner together provide optical magnification of up to 150x,
depending on the zoom setting of the video lens, on the computer
monitor.
The video system illuminator is a separate box (
to the video system by a fiber light pipe. The light pipe can be separated
from the illuminator and/or optics by loosening the knobs at either end.
Turn on the illuminator with the switch on the front of the box. Increase
intensity by turning the Illumination Level knob clockwise, or decrease
it by turning it counterclockwise.
Figure 4-29
Turn on the illuminator and adjust intensity
NOTE
The video system camera automatically adjusts to the brightness level.
Above a certain level turning up the illumination will no longer have
any effect, as the camera will compensate to counteract the increased
intensity.
NOTE
For critical measurements use the video system to align the sample, then
turn off the illuminator to minimize system noise, as the fan in the light
source can transmit vibrations to the microscope.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...