Preparing for Imaging
5
Keysight 5500 SPM User’s Guide
5-26
Mounting the Sample
The Keysight 5500 SPM accepts a wide variety of sample plates,
including specialized plates for imaging in liquid, in controlled
temperature, etc. To use a sample plate:
1
Mount the sample to the sample plate.
In general, samples should be held in place securely enough to
prevent drift or creep during measurement, but not so firmly as to
induce stress in the sample. Several mounting methods are available.
A common approach is to mount the sample on a magnetically
attractive backing which can then be held by the magnet on the
standard sample plate. Large, flat samples can be held down using
the clips from the liquid cell plate. Double-back tape can also be
used, though the tape tends to deform easily and can lead to creep
during imaging.
2
Place the front alignment tab of the sample plate over the front
alignment pin, as in image A of
CAUTION
Verify that there is enough space between the scanner and sample plate
such that the tip will not contact the sample plate once the plate is
mounted. Contact with the plate will certainly damage the probe and
perhaps the sample.
It is recommended that the scanner be moved up 100 microns or more
whenever changing a probe or plate to avoid damage. Simply hold the
switch on the front of the HEB in the “Open” for 5-10 seconds to
provide adequate clearance.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...