53210A Software Installation and Interface Configuration
72
Keysight 53210A User’s Guide
Firmware and Driver Updates
Firmware and driver updates (when available) for the 53210A are available via the
Web. This section contains information for locating and downloading the updates
to your computer, and then installing the updates in the instrument.
Disabling Calibration Security
Before firmware updates can be installed, the counter’s calibration security
feature must be
disabled
. This done using the command:
CALibration:SECurity:STATe {OFF|ON}, <code>
CALibration:SECurity:STATe?
(query form)
To unsecure the instrument, you must provide the security code in addition to
setting the state to
OFF
. The security code is set to
KT53210A
when the instrument
is shipped from the factory. Note, however, the security code may have been
changed once the counter has been in use.
After the firmware update is complete, the counter can be re-secured by setting
the state to
ON
and providing the security code. The calibration count is
unaffected
when changing the security state or by a firmware update.
The calibration count can be read with the command:
CALibration:COUNt?
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...