Example: Limit Checking
The following is an example of enabling and using limit checking.
CONF:FREQ (@1)
// configure measurement
SAMP:COUN 500
// check limit on 500 rdgs
CALC:STAT ON
// enable math operations
CALC:LIM:STAT ON
// enable limit checking
CALC:LIM:LOW 99.9E3;UPP 100.1E3 // set limits
INIT
// initiate the measurements
*WAI
// wait for readings to complete
STAT:QUES:EVEN?
// query questionable data register
Notice that the limits are set on the same line to prevent a settings conflict error.
This Error can also be avoided by enabling limit checking
after
the limits are set.
Clearing Limit Conditions
The ‘Limit’ annunciator is turned off and bits 11 and 12 (only) in the Questionable
Data Register are cleared by any of the following:
– enabling limit checking
- CALCulate[1]:LIMit[:STATe] ON
– a new measurement cycle
- INITiate:IMMediate
,
READ?
,
MEASure?
– sending a new SCPI command or changing a current SCPI parameter
– a reset or instrument preset
-
*RST, SYSTem:PRESet
These steps also clear
all readings
in reading memory. (
*RST
and
SYSTem:PREset
additionally reset the lower and upper limits to 0.0.)
To turn off the ‘Limit’ annunciator and clear only the limit-sensing bits (11/12 in
the Condition register)
without
clearing reading memory, the following command
is used:
CALCulate[1]:LIMit:CLEar[:IMMediate]
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...