53210A Math, Graphing, and Data Logging
6
Keysight 53210A User’s Guide
151
Using the Scale Functions
The counter scaling functions include null, percent change (
PCT
), part per million
change (
PPM
), part per billion change (
PPB
), and scale (
Mx-B
). The function is
selected with the command:
CALCulate[1]:SCALe:FUNCtion {NULL|PCT|PPM|PPB|SCALe}
CALCulate[1]:SCALe:FUNCtion?
(query form)
NULL
- performs a null operation. The result is the measurement minus the
reference value. (See “
” for information on setting the
reference).
PCT
- performs a percent change operation. The result is the difference between
the measurement and the reference value, expressed in percentage.
PPM
- performs a parts per million change operation. The result is the difference
between the measurement and the reference value, expressed in parts per million.
PPB
- performs a parts per billion change operation. The result is the difference
between the measurement and the reference value, expressed in parts per billion.
SCALe
- performs an Mx-B operation. The result is the measurement (x) multiplied
by the gain value M (
CALCulate1:SCALe:GAIN
command) minus the offset value B
(
CALCulate1:SCALe:OFFSet
). If
CALCulate1:SCALe:INVert
is enabled (
On
), the
measurement will first be inverted (1/x), resulting in a M/x-B operation.
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...