3
Programming
214
Keysight 53147A/148A/149A Operating and Programming Guide
:DISPlay Subsystem
This subsystem controls the selection and presentation of textual information on
the instrument's display. This information includes measurement results. :
DISPlay
is independent of, and does not modify, how data is returned to the controller.
See the section titled
“Programming the Instrument to Display Results”
on
COMMAND
:DISPlay[:WINDow]:BACKground[:STATe] . . . <Boolean>
Turns the display backlight ON or OFF.
QUERY
RESPONSE
– Single ASCII-encoded byte, 0 or 1.
– A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENTS
–
*RST: ON
– Turning the display backlight OFF conserves battery power (if the instrument
has the Battery option).
– If the instrument is operating from battery power (Battery option required), the
display backlight is automatically turned off after a period of inactivity. When
activity is subsequently detected (a measurable signal is applied, a front-panel
key is pressed, or a setting is changed by a command sent over the GPIB or the
RS-232 interface), the display backlight is restored to its previous state (i.e., if
the backlight was set to
OFF
, it remains off; if the backlight was set to
ON
, it is
turned on).
RELATED
FRONT-PANEL
KEYS
Shift + On/Off
COMMAND
:DISPlay:ENABle . . . <Boolean>
Sets or queries whether the entire display (annunciators and indicators, with the
exception of
Rmt
) is visible.
QUERY
RESPONSE
– Single ASCII-encoded byte, 0 or 1.
– A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENTS
– *RST: ON