94
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #17: Decoding, Triggering, and Searching on RS232/UART Serial
Bus Signals
In this lab you will learn how to set up the scope to decode and trigger on transmit
and receive RS232/UART serial bus traffic. In addition, you will learn how to use
the scope’s automatic RS232/UART Search & Navigation capability, as well as
Segmented Memory acquisition. This lab does not provide a tutorial on the
RS232/UART protocols and signaling. To complete this lab, your scope must be
licensed with the RS232/UART trigger and decode option (Option COMP). You can
verify the installed options on your oscilloscope at
[Help] > About Oscilloscope
.
1
Connect the channel-1 probe to the Demo 1 terminal and ground.
2
Connect the channel-2 probe to the Demo 2 terminal and ground.
3
Press
[Default Setup]
on the scope’s front panel.
4
Press
[Help]
; then tap the
Training Signals
softkey.
5
Using the Entry knob, select the
RS232/UART
signal; then tap the
Output
softkey
to turn it on.
6
Set channel-1 to
1.00 V/div
.
7
Set channel-1’s offset/position to
-1 V
.
8
Press the
[2]
front panel key to turn on channel-2.
9
Set channel-2 to
1.00 V/div
.
10
Set channel-2’s offset/position to
+2.3 V
.
11
Push the trigger level knob to automatically set the trigger level to
approximately
50%
.
12
Set the scope’s timebase to
2.00 ms/div
.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...