80
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #15: Decoding, Triggering, and Searching on I
2
C Serial Bus
Signals
In this lab you will learn how to set up the scope to decode and trigger on I
2
C
serial bus traffic. In addition, you will learn how to use the scope’s automatic I
2
C
Search & Navigation capability, as well as use Segmented Memory acquisition to
capture multiple and consecutive occurrences of a particular Read operation. This
lab does not provide a tutorial on the I
2
C protocol and signaling. To complete this
lab, your scope must be licensed with the I
2
C trigger and decode option (Option
EMBD). You can verify the installed options on your oscilloscope at
[Help] > About
Oscilloscope
.
1
Connect the channel-1 probe to the Demo 1 terminal and ground.
2
Connect the channel-2 probe to the Demo 2 terminal and ground.
3
Press
[Default Setup]
on the scope’s front panel.
4
Press
[Help]
; then tap the
Training Signals
softkey.
5
Select and enable the
I
2
C
signal.
6
Set channel-1 to
1.00 V/div
.
7
Press the
[2]
front panel key to turn on channel-2.
8
Set channel-2 to
1.00 V/div
.
9
Set channel-2’s offset/position to
+3.50 V
.
10
Push the trigger level knob to set the trigger level to approximately
50%
.
11
Set the scope’s timebase to
500.0 µs/div
.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...