Oscilloscope Familiarization Labs
2
4000 X-Series Oscilloscopes Advanced Training Guide
39
Lab #6: Using Mask Test
With mask testing you can set up a pass/fail test criteria for automatically testing
waveforms to see if they conform to specific wave shapes. In this lab we will test a
digital signal that includes an infrequent glitch. With InfiniiVision’s
hardware-based mask testing capability, we will be able to test over 200,000
waveforms per second and gain insight into the statistical occurrences of this
particular glitch. To enable Mask Testing, your oscilloscope must have the mask
test option installed (Option DSOX4MASK). You can verify the installed options on
your oscilloscope at
[Help] > About Oscilloscope
.
1
Connect the channel-1 probe to the Demo 1 terminal and ground.
2
Press
[Default Setup]
on the scope’s front panel.
3
Press
[Help]
; then press the
Training Signals
softkey.
4
Turn on
Clock with Infrequent Glitch
signal.
5
Set channel-1 to
500 mV/div
and position/offset to
1.0 V
.
6
Push the Trigger Level knob to set the trigger level at
50%
.
7
Set the timebase to
20 ns/div.
8
Set the waveform intensity to
100%
by first pressing the
[Intensity]
button; then
turn the Entry knob.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...