14
4000 X-Series Oscilloscopes Advanced Training Guide
2
Oscilloscope Familiarization Labs
Lab #1: Using Cursors and Automatic Parametric Measurements
During this first hands-on lab, you will learn how to make simple voltage and
timing measurements using the scope’s manually positioned measurement
cursors, as well as perform similar measurements using the scope’s automatic
parametric measurement capability.
1
Connect the channel-1 probe to the Demo 1 terminal and ground.
2
Press
[Default Setup]
on the scope’s front panel.
3
Press
[Help]
; then press the
Training Signals
softkey.
4
Using the touch screen
Training Signals
menu, select the
Repetitive Pulse with
Ringing
signal; press the
Output
softkey to turn it on.
5
Set channel-1’s V/div setting to
500 mV/div
.
6
Set channel-1’s position/offset to
1.40 V
.
7
Push the trigger level knob to automatically set the trigger level at
approximately
50%
.
8
Set the scope’s timebase to
500.0 ns/div
.
Figure 1
Oscilloscope setup to capture and display a repetitive digital pulse with ringing
and overshoot.