Frequency Accuracy Calibration
This calibration procedure applies to all models (34460A/61A/65A/70A).
Review
and
Gain Calibration Considerations
before beginning this procedure.
Configuration: Frequency, 10 V range
The frequency accuracy of the Fluke 5720A is insufficient to calibrate the DMM. Its frequency output needs
to be calibrated against a more accurate reference. The Keysight 33500B Series Function/Arbitrary Wave-
form Generator is recommended for this calibration procedure.
1.
Configure the instrument for frequency measurements at 10 V and 10 kHz, with 1 second gate time.
2.
Apply a 10 kHz sine wave between 9 and 11 Vrms.
3.
Enter the actual applied input frequency (see
). A Calibration Step Suc-
ceeded message indicates success; if the display shows Calibration Step Failed, check the input value,
range, function, and entered calibration value and repeat the calibration step.
4.
Store the new calibration constants (see
Storing the Calibration Constants
).
This calibration takes approximately 1 second to complete.
Keysight True
volt
Series Operating and Service Guide
547
Calibration Adjustment Procedures
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...