Frequency Accuracy Verification
Configuration: Frequency (CONFigure:FREQuency DEF, MIN)
1.
Make sure you have read
.
2.
Select the Frequency function, default range, 3 Hz filter, and 1 second aperture.
3.
Connect the Keysight 33500B Series Function/Arbitrary Waveform Generator to the input terminals.
For the 34461A/65A/70A, use the front panel input terminals and select the front input terminals with
the
Front/Rear
switch.
4.
Select each range in the order shown below. Provide the indicated input voltage and frequency. Com-
pare measurement results to the test limits shown below. (Be certain to allow for appropriate source
settling.)
Values for 34460A DMM
Input
Error from Nominal
Vrms
Frequency
Range
Quick
Check
24 hours 90 days
1 year
2 years
0.1 Vrms
10 Hz
1 Vrms
± 3 mHz
± 3.5 mHz
0.01 Vrms
300 kHz
0.1 Vrms
Q
± 60 Hz
± 240 Hz ± 360 Hz ± 510 Hz
Values for 34461A DMM
Input
Error from Nominal
Vrms
Frequency
Range
Quick
Check
24 hours 90 days
1 year
2 years
0.1 Vrms
10 Hz
1 Vrms
± 3 mHz
± 3.5 mHz
0.01 Vrms
300 kHz
0.1 Vrms
Q
± 60 Hz
± 180 Hz ± 300 Hz ± 450 Hz
Values for 34465A and 34470A DMMs
Input
Error from Nominal
Vrms
Frequency
Range
Quick
Check
24 hours 90 days
1 year
2 years
0.1 Vrms
10 Hz
1 Vrms
± 3 mHz
0.01 Vrms
300 kHz
0.1 Vrms
Q
± 60 Hz
± 150 Hz ± 210 Hz ± 270 Hz
Q
:
Quick performance verification test point.
Keysight True
volt
Series Operating and Service Guide
515
Performance Verification
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...