SYSTem:COMMunicate:GPIB:ADDRess <
address
>
SYSTem:COMMunicate:GPIB:ADDRess?
Assigns instrument's GPIB (IEEE-488) address, which is displayed at power-on. Each device on the GPIB
interface must have a unique address.
The GPIB interface is an optional feature. See
for details.
Parameter
Typical Return
0 to 30, default 22 +15
Set GPIB address to 15:
SYST:COMM:GPIB:ADDR 15
l
Your computer's GPIB interface card has its own address. Avoid using this address for any instrument
on the GPIB bus.
l
Cycle power for this command to take effect.
l
GPIB is enabled or disabled by
.
l
The GPIB interface is an optional feature. See
for details.
l
This setting is non-volatile; it is not changed by power cycling, a Factory Reset (
), or an Instru-
ment Preset (
).
l
This parameter is set to its default value when the instrument is shipped from the factory and after
.
Keysight True
volt
Series Operating and Service Guide
407
SCPI Programming Reference
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...