SYSTem:DATE <
year
>, <
month
>, <
day
>
SYSTem:DATE?
Sets the date for the instrument's real-time clock. Use
to set the time.
Parameter
Typical Return
<
year
2000 to 2099
<
month
> 1 to 12
<
day
> 1 to 31
+2011,+07,+26
Set system date to July 26, 2011:
SYST:DATE 2011, 7,26
l
The real-time clock is used for Mass Memory (MMEMory) system file time stamps.
l
The real-time clock is battery-backed and retains the date and time when power is off.
l
The date and time are not affected by
or
l
The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set
the date and time when you first receive your instrument.
SCPI Programming Reference
396
Keysight True
volt
Series Operating and Service Guide
Содержание 34460A
Страница 1: ...Operating and Service Guide Keysight Truevolt Series Digital Multimeters ...
Страница 183: ...Measurement Tutorial 184 Keysight Truevolt Series Operating and Service Guide ...
Страница 385: ...STATus Subsystem SCPI Programming Reference 386 Keysight Truevolt Series Operating and Service Guide ...
Страница 541: ...Next Procedure Calibration Adjustment Procedures 542 Keysight Truevolt Series Operating and Service Guide ...
Страница 556: ...Next Procedure Keysight Truevolt Series Operating and Service Guide 557 Calibration Adjustment Procedures ...