[SENSe:]{FRESistance|RESistance}:SECondary {"OFF"|"CALCulate:DATA"}
[SENSe:]{FRESistance|RESistance}:SECondary?
Selects the secondary measurement function for all resistance measurements.
Parameter
Typical Return
{"OFF"|"CALCulate:DATA"}, default "OFF"
"CALC:DATA"
Configure resistance and secondary measurements. Make resistance measurement and retrieve secondary
measurement.
CONF:RES AUTO,MAX
RES:SEC "CALC:DATA"
READ?;DATA2?
Typical Response: +3.01566373E-10;,+9.913
l
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
math operations are done (including NULL).
l
Use
or
to start the measurement. Use
to retrieve the secondary meas-
urement.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
356
Keysight True
volt
Series Operating and Service Guide